SPECIFICATIONS
|
MODEL
|
7151
|
7152
|
7160
|
7162
|
AC WITHSTAND VOLTAGE
|
Output Rating
|
5KVAC/12mA
|
|
Range
|
Resolution
|
Accuracy
|
Output Voltage, KVAC
|
0-5.00
|
0.01
|
± ( 2% of setting + 5V )
|
Output Frequency
|
50Hz/60Hz ±100ppm, User Selection
|
Output Waveform
|
Sine Wave ,THD.<2% (Resistive Load), Crest Factor = 1.3 - 1.5
|
Output Regulation
|
± (1% of output + 5V), From no load to full load
|
SETTINGS
|
Hi-Limit AC Current, mA
|
0.10 - 12.00
|
0.01
|
± ( 2% of setting + 2counts )
|
Lo-Limit AC Current, mA
|
0.00 - 12.00
|
0.01
|
± ( 2% of setting + 2counts )
|
Ramp Time, second
|
0.1 - 999.9
|
0.1
|
± ( 0.1% + 0.05sec )
|
Dwell Time, second
|
0, 0.2 - 999.9
|
0.1
|
(0=continuous)
|
Arc Detection
|
0, 1 - 9 ranges (0=OFF, 9 is the most sensitivity)
|
DC WITHSTAND VOLTAGE (7120,7122 ONLY)
|
Output Rating
|
6KVDC/5mA
|
Output Voltage, KVDC
|
0 - 6.00
|
0.01
|
± ( 2% of setting + 5V )
|
Output Ripple
|
<5% ( 6KV/5mA at Resistive Load )
|
SETTINGS
|
Hi-Limit DC Current, mA
|
0.02 - 5.00
|
0.01
|
± ( 2% of setting + 2counts )
|
Lo-Limit DC Current, mA
|
0.00 - 5.00
|
0.01
|
± ( 2% of setting + 2counts )
|
Ramp Time, second
|
0.1 - 999.9
|
0.1
|
± ( 0.1% + 0.05sec )
|
Dwell Time, second
|
0, 0.2 - 999.9
|
0.1
|
(0=continuous)
|
Arc Detection
|
0, 1 - 9 ranges (0=OFF, 9 is the most sensitivity)
|
Discharge Time
|
< 200 msec
|
INSULATION RESISTANCE (7112,7122 ONLY)
|
Output Rating
|
1KVDC/1000MΩ
|
Output Voltage, KVDC
|
0.10 - 1.00
|
0.01
|
± ( 2% of setting + 3V )
|
SETTINGS
|
Hi-Limit Resistance, MΩ
|
0, 1 - 1000 (0=OFF)
|
1
|
± ( 3% of setting+2counts ) at > 500VDC
|
Lo-Limit Resistance, MΩ
|
1 - 1000
|
1
|
± ( 7% of setting +2counts ) at <500VDC
|
Delay Time, second
|
0, 0.5 - 999.9
|
0.1
|
± ( 0.1% + 0.05sec )
|
(0=continuous)
|
MEASUREMENT
|
AC/DC Voltage, KV
|
0 - 6.00
|
0.01
|
± ( 1.5% of reading + 1count )
|
DC Voltage, KV (IR only)
|
0.10 - 1.00
|
0.01
|
± ( 1.5% of reading + 3V )
|
AC Current, mA
|
0.10 - 12.00
|
0.01
|
± ( 2% of reading + 2counts )
|
DC Current, mA
|
0.02 - 5.00
|
0.01
|
± ( 2% of reading + 2counts )
|
Resistance, MΩ
|
1 - 1000
|
0.001
|
± ( 3% of reading+2counts ) at > 500VDC
|
(Auto Range)
|
0.01
|
± ( 7% of reading +2counts ) at <500VDC
|
|
0.1
|
|
GENERAL
|
Input Voltage AC
|
115/230Vac±15%, 50/60Hz ± 5%, Fuse 3.15A Fast
|
PLC Remote Control
|
Input : Test , Reset
|
Output : Pass , Fail , Processing
|
Memory
|
5 Sets Store/Recall of each Memory, 4 Test Mode Selectable ( W, I, W ~ I, I ~ W Only 7112, 7122 )
|
Display
|
16×2 LCD with back light
|
Key Lock
|
To prevent unauthorized alteration of the test Parameters
|
Calibration
|
Build-in software and external calibrated meters
|
Environment
|
0-40℃, 20-80%RH
|
Dimension/Net Weight
|
280mm (W)×89mm(H)×370mm(D)/ 9Kg
|
STANDARD ACCESSORIES
|
Power Cord(10A)
|
×1
|
Fuses
|
×3 (Including a spare contained in the fuse holder)
|
Hipot Test Lead ,1.5m(1101)
|
×1
|
Hipot Return Lead ,1.8m(1102)
|
×1
|
*product specifications are subject to change without notice.
|